WILKS ENTERPRISE Inc.
Thursday, 11 March 2010

Oil/Grease AnalyzersBiofuel AnalyzersSoot AnalyzerInfraSpec VFA-IR SpectrometersInfraRan Specific Vapor AnalyzersInfraCal Filtometers
Request More Information
Oil/Grease Analyzers
PDF Print E-mail
 


 Model CVHModel HATR-T2
Model CH

Minimum Detection Limit
(water)

2 ppm
8 ppm
40 ppm
Minimum Detection Limit
(soil)
10 ppm
40 ppm
200 ppm
Solvent
Perchloroethylene, Freon-113, S-316
Hexane, Pentane, Vertrel MCA
Hexane, Pentane, Vertrel MCA
Component Measured
Measures all hydrocarbons Measures hydrocarbons with a boiling point higher than the solvent - volatile hydrocarbons will be evaporated along with the solvent.
Measures hydrocarbons with a boiling point higher than the solvent - volatile hydrocarbons will be evaporated along with the solvent.
Instrument Repeatability
+/- 1 ppm
+/- 2 ppm
+/- 4 ppm
OtherSolvents are heavier than water and require a separatory funnel or inverted vial with a septa cap to collect a solvent sample.
Solvents are lighter than water and solvent sample can be collected from the top of the container.
Solvents are lighter than water and solvent sample can be collected from the top of the containe

Over 2,000 Wilks Oil/Grease Analyzers in World-Wide Use Today!                          

The InfraCal TOG/TPH Analyzers provide an easy and accurate field-proven infrared analysis method for on-site measurement of total oil and grease (TOG) and total petroleum hydrocarbon (TPH) levels in water and soil samples as well as, fats, oil and grease (FOG) levels in water. With well over 2,000 InfraCal TOG/TPH Analyzers in worldwide operation, they have become the industry standard for TOG and TPH measurements. Depending on sensitivity requirements and the extracting solvent being used, several InfraCal models are available - cuvette holder (CVH), horizontal platform (HATR-T2) and card reader (CH).

  • Fast - TOG/TPH or FOG determinations in under 10 - 15 minutes, including extraction
  • Accurate - Detect levels as low as 2 ppm or as high as 5000+ ppm (depending on model selected)
  • Portable/Rugged -- 6.5 x 6.5 x 5 inches, can be battery operated, ideal for field or laboratory applications
  • Lightweight -- less than 5 lbs
  • Easy to use -- no knowledge of infrared techniques needed
  • Direct readout in desired units
  • No moving parts
  • Versatile  -- Models usable with hexane, pentane, Vertrel MCA, perchloroethylene, Freon-113, AK-225 or S-316 extraction procedures
  • Measures both Aliphatic and Aromatic components in sample
  • RS-232 connection for data transmission to a PC or printer

A range of optional accessories expands the capabilities and versatility of the InfraCal TOG/TPH Analyzers. The battery pack that provides up to 16 hours of continuous operation or the cigarette lighter adapter plug enables the analyzer to be used for field testing. A small, portable printer provides time stamped printouts of analytical results. Transporting the InfraCal TOG/TPH Analyzer and various accessories is easy with our sturdy carrying case with pre-diced pluck foam insert.

Whether your application is....